Measurements and Result Displays
R&S
®
ESW
133
User Manual 1177.6298.02 ─ 08
"Transducer"
Draws a trace that shows the correction values of all active trans-
ducer factors in the currently selected frequency range. In case trans-
ducer factors overlap each other, the correction values are aggrega-
ted.
Available for Trace 1 and in the receiver application only.
Note
: When you select the "Transducer" trace, the application tempo-
rarily removes all other traces from the diagram. You can restore the
scan traces by selecting the original trace mode of trace 1 again.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE
View transducer:
[SENSe:]CORRection:TRANsducer:VIEW
Detector
Selects the trace detector.
Note:
In the receiver application, you can apply different detectors to the bargraph, the
scan and the final measurement.
"Max Peak"
detector.
"Min Peak"
detector.
"RMS"
"Average"
"Quasipeak"
"CISPR Aver-
age"
"RMS Aver-
age"
detector.
"None"
Ignores the corresponding trace during the final measurement.
Available for the final measurement.
Remote command:
Bargraph:
[SENSe:]DETector<t>:RECeiver[:FUNCtion]
Scan:
[SENSe:]DETector<t>[:FUNCtion]
Final measurement:
[SENSe:]DETector<t>:FMEasurement
Interactive
Turns interactive measurements on and off.
For more information, see
"Automatic vs interactive final measurements"
Remote command:
9.3.5.2
Working with the Final Measurement Results
Access
: "Overview" > "Test Automation" > "Final Result"
When the final measurement is done, the R&S
ESW writes the results into a table that
looks and feels similar to the peak list. Editing entries is, however, not possible.
Test Automation