![Rohde & Schwarz ESW Series User Manual Download Page 396](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816396.webp)
Network and Remote Operation
R&S
®
ESW
396
User Manual 1177.6298.02 ─ 08
Language
Comment
8566B
8568A
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568A_DC
Uses DC input coupling by default if supported by the instrument
8568B
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568B_DC
Uses DC input coupling by default if supported by the instrument
8591E
Compatible to 8594E
8594E
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
PSA89600
PSA
Notes:
●
If you select a language other than "SCPI", the GPIB address is set to 18 if it was
20 before.
●
The Start/stop frequency, reference level and number of sweep points are adapted
to the selected instrument model.
●
When you switch between remote control languages, the following settings or
changes are made:
SCPI:
The instrument performs a PRESET.
8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM; FSEK:
–
The instrument performs a PRESET.
–
The following instrument settings are changed:
Table 14-21: Instrument settings for emulation of 8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM;
FSEK instruments
Model
# of Trace
Points
Start Freq.
Stop Freq.
Ref Level
Input Coupling
8566A/B
1001
2 GHz
22 GHz
0 dBm
AC
8568A/B
1001
0 Hz
1.5 GHz
0 dBm
AC
8560E
601
0 Hz
2.9 GHz
0 dBm
AC
8561E
601
0 Hz
6.5 GHz
0 dBm
AC
8562E
601
0 Hz
13.2 GHz
0 dBm
AC
8563E
601
0 Hz
26.5 GHz
0 dBm
AC
8564E
601
0 Hz
40 GHz
0 dBm
AC
8565E
601
0 Hz
50 GHz
0 dBm
AC
8594E
401
0 Hz
3 GHz
0 dBm
AC