Measurements and Result Displays
R&S
®
ESW
136
User Manual 1177.6298.02 ─ 08
............................................................................................ 137
LISN Type
Selects the LISN used for the measurement.
The following LISNs are supported by the R&S
ESW:
●
R&S ENV216
●
R&S ENV432
●
R&S ENV4200
●
R&S ESH2-Z5
●
R&S ESH3-Z5
Select "Off" when you are performing measurements without a LISN.
Remote command:
Phase
Selects the LISN phase to be measured.
Phases L2 and L3 are available when you are using the R&S
ENV4200, R&S
432 or
R&S
ESH2-Z5 (four-line networks).
In the spectrum application, the I/Q analyzer and the analog demodulator, you can
measure on one phase at a time (N, L1, L2
or
L3).
In the receiver application, phase selection works as follows.
●
You can measure more than one phase in a single measurement (N, L1, L2
and /
or
L3).
Instead of selecting the phase in the "Phase" column of the dialog, select the pha-
ses you want to measure in the "Scan" or the "Final Test" column ("On" considers
the phase in the measurement, "Off" ignores the phase in the measurement).
If you are measuring more than one phase, the R&S
ESW performs a measure-
ment for each of the selected phases.
●
You can select different phase (combinations) for the scan and the final measure-
ment.
The selection in the "Scan" and "Final Test" blocks of the dialog box override the
selection in the "Phase" block of the dialog.
Example:
•
Phase L1 is selected in the "Phase" block.
•
Phase L1 and N are selected in the "Scan" block.
•
Phase L1 is off in the "Final Test" block, but phase N is on.
➙
During the scan, the R&S
ESW measures both selected phases.
➙
During the final measurement, the R&S
ESW measures phase N, but not phase
L1.
Test Automation