Remote Commands in the Receiver Application
R&S
®
ESW
565
User Manual 1177.6298.02 ─ 08
CALCulate<n>:LIMit<li>:LOWer:OFFSet
<Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
<li>
Parameters:
<Offset>
Numeric value.
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:STATe
<State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
CALCulate<n>:LIMit<li>:STATe
<State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use
.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
Suffix:
<n>
.
irrelevant
Analysis