![Rohde & Schwarz ESW Series User Manual Download Page 544](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816544.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
544
User Manual 1177.6298.02 ─ 08
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
<SearchLimit>
This command defines the left limit of the marker search range for
all
markers in
all
windows.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<SearchLimit>
The value range depends on the frequency range or measure-
ment time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
left diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
"Search Limits (Left / Right)"
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range for
all
markers in
all
windows.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
"Search Limits (Left / Right)"
CALCulate<n>:THReshold
<Level>
This command defines a threshold level for the marker peak search (for
all
markers in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<Level>
Numeric value. The value range and unit are variable.
*RST:
-120 dBm
Default unit: DBM
Analysis