Remote Commands in the Receiver Application
R&S
®
ESW
482
User Manual 1177.6298.02 ─ 08
15.5.12
CISPR APD Measurement Configuration
The remote commands to configure CISPR APD measurements are similar to those
available for the APD measurement in the Spectrum application.
For a comprehensive description, refer to the user manual of the Spectrum application.
CALCulate<n>:MARKer<m>:Y:PERCent
.......................................................................... 482
CALCulate<n>:MARKer<m>:Y:PERCent
<Probability>
This command sets a marker to a particular probability value. You can query the corre-
sponding level with
.
Using the command turns delta markers into normal markers.
This command is available for CCDF measurements.
Suffix:
<n>
.
<m>
Parameters:
<Probability>
Range:
0 % to 100 %
Default unit: %
Example:
CALC1:MARK:Y:PERC 95PCT
Positions marker 1 to a probability of 95 %.
Manual operation:
See
15.5.13
Programming Example: Performing a Sequence of Measurements
This example demonstrates how to perform several measurements in a sequence in a
remote environment.
//2xSpectruman 2xIQ, start Sequencer at the end, test OPC?
// ------------------------------------------------------------------------
//------Preparing the instrument and first channel -----------
*RST
//Activate new IQ channel
INSTrument:CREate:NEW IQ,'IQ 1'
//Set sweep count for new IQ channel
SENS:SWEEP:COUNT 6
//Change trace modes for IQ channel
DISP:TRAC1:MODE BLANK
DISP:TRAC2:MODE MAXH
DISP:TRAC3:MODE MINH
//Switch to single sweep mode
INIT:CONT OFF
Measurements and Result Displays