Remote Commands in the Receiver Application
R&S
®
ESW
561
User Manual 1177.6298.02 ─ 08
CALCulate<n>:DLINe<dl>:STATe
<State>
This command turns a display line on and off
Suffix:
<n>
.
<dl>
1 | 2
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DLIN2:STAT ON
Turns on display line 2.
CALCulate<n>:FLINe<dl>
<Frequency>
This command defines the position of a frequency line.
Suffix:
<n>
.
<dl>
1 | 2
frequency line
Parameters:
<Frequency>
Note that you can not set a frequency line to a position that is
outside the current span.
Range:
0 Hz to Fmax
*RST:
(STATe to OFF)
Default unit: HZ
Example:
CALC:FLIN2 120MHz
Sets frequency line 2 to a frequency of 120 MHz.
Manual operation:
See
CALCulate<n>:FLINe<dl>:STATe
<State>
This command turns a frequency line on and off
Suffix:
<n>
.
<dl>
1 | 2
frequency line
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
Analysis