Remote Commands in the Receiver Application
R&S
®
ESW
452
User Manual 1177.6298.02 ─ 08
Suffix:
<t>
.
Parameters:
<Detector>
You can select up to four detectors, one for each active bar-
graph:
<Detector>, [<Detector>, <Detector>, <Detector>]
AVERage
Selects the Average detector.
CAVerage
Selects the CISPR Average detector.
CRMS
Selects the CISPR RMS detector.
NEGative
Selects the Min Peak detector.
POSitive
Selects the Max Peak detector.
QPEak
Selects the Quasipeak detector.
RMS
Selects the RMS detector.
*RST:
AVERage
Example:
//Select bargraph detector
DET:REC POS,AVER,RMS
Manual operation:
See
[SENSe:]SWEep:TIME
<Time>
This command defines the measurement or acquisition time for bargraph measure-
ments.
For scans not based on a scan table, the command also defines the measurement
time for the scan.
Parameters:
<Time>
<numeric value>
Range:
5 µs to 100 s
*RST:
0.1
Default unit: s
Example:
//Define measurement time
SWE:TIME 10s
Manual operation:
See
Measurements and Result Displays