This setting is only available if a HP language is selected (see
Remote command:
Wideband
This setting defines which option is returned when the
*OPT?
query is executed,
depending on the state of the wideband option.
It is only available for PSA89600 emulation.
"Off"
No wideband is used.
The option is indicated as "B7J".
"40 MHz"
The 40
MHz wideband is used.
The option is indicated as "B7J, 140".
"80 MHz"
The 80
MHz wideband is used.
The option is indicated as "B7J, 122".
Remote command:
Revision String
Defines the response to the
REV?
query for the revision number.
(HP emulation only, see
Max. 36 characters are allowed.
Remote command:
Resetting the Factory Revision
Resets the response to the
REV?
query for the revision number to the factory default
(HP emulation only, see
Remote command:
14.4.4
LXI Settings
Access
: [SETUP] > "N Remote" > "LXI" tab
On the R&S
ESW the LXI Class C functionality is already installed and enabled; thus,
the instrument can be accessed via any web browser (e.g. the Microsoft Internet
Explorer) to perform the following tasks:
●
modifying network configurations
●
modifying device configurations
●
monitoring connections from the device to other devices
The "LXI" tab of the "N Remote" dialog box provides basic LXI functions and
information for the R&S
ESW.
Alternatively, you can change the LAN settings using the LXI Web browser interface.