Limit check for the corresponding frequency has failed. This indicates a potential
interferer.
●
Rows in blue highlighting
Indicates the currently selected row.
●
Trace / Detector
Shows the number of the trace and the detector used for that trace.
●
Frequency
Shows the frequency where the peak was found.
●
Level
Shows the absolute level value of the peak. The unit depends on the unit you have
selected.
The result also contains information about the phase that was measured when you
are working with a LISN (L1, L2, L3, N).
●
Delta Limit
Shows the distance of the peak to the nearest limit line that has been assigned to
the corresponding trace. The distance is a value in dB. Available if you are using
limit lines.
●
Comment
Allows you to add a comment to each result as required.
9.3.6
Configuring Line Impedance Stabilization Networks (LISN)
Access
: "Overview" > "Test Automation" > "LISN"
The R&S
ESW supports several LISN models and provides functionality to control
these devices.
For more information about using a LISN, see
Chapter 9.3.1.3, "Line Impedance Stabi-
lization Network (LISN) Control"
Figure 9-16: Overview of the "LISN" configuration
Note: the "Scan" and "Final Test" blocks are available in the Receiver application.
The remote commands required to configure the LISN are described in
ter 15.5.11, "LISN Configuration"
Test Automation