User Manual 1177.6298.02 ─ 08
Generator Type
Selects the generator type and thus defines the generator setup file to use.
For an overview of supported generators, see
"Overview of Supported Generators"
on page 154. For information on generator setup files, see
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE
Interface
Type of interface connection used.
For details on which signal generators support which interfaces, see the documenta-
tion of the corresponding signal generator.
●
GPIB
●
TCP/IP
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface
TTL Handshake
If available for the specified generator type, this option activates TTL synchronization
via handshake.
Using the TTL interface allows for considerably higher measurement rates, because
the frequency stepping of the R&S
ESW is directly coupled with the frequency stepping
of the generator.
For more information on TTL synchronization, see
For an overview of which generators support TTL synchronization see
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK
GPIB Address/TCPIP Address / Computer Name
For LAN connections: TCP/IP address of the signal generator
For GPIB connections: GPIB address of the signal generator.
Remote command:
SYSTem:COMMunicate:GPIB:RDEVice:GENerator<gen>:ADDRess
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess
Reference
Selects the internal R&S
ESW or an external frequency reference to synchronize the
R&S
ESW with the generator (default: internal).
Remote command:
SOURce<si>:EXTernal<gen>:ROSCillator[:SOURce]
Edit Generator Setup File
Displays the setup file for the currently selected
an editor.