![Rohde & Schwarz ESW Series User Manual Download Page 131](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816131.webp)
Measurements and Result Displays
R&S
®
ESW
131
User Manual 1177.6298.02 ─ 08
Availability of measurement parameters
When you interrupt the scan or the final measurement, you can change only parame-
ters that have an immediate effect on the measurement. All other parameters (e.g. trig-
ger settings) are unavailable.
Sequence for an interactive measurement
If you select the interactive final measurement, the R&S
ESW initiates the following
sequence.
1. The R&S
ESW tunes to the first frequency in the peak list. The measurement con-
figuration is as defined previously.
2. The R&S
ESW positions a marker on that frequency and interrupts the measure-
ment.
3. While the measurement is interrupted, you can change any setting that is available.
In addition, you have several options on how to proceed.
● Skip the current frequency.
Positions the marker on the next frequency in the peak list without performing a
final measurement on the current frequency.
● Get max hold result for the current frequency.
Writes the highest level that was measured for that frequency during the scan
to the final peak list without performing a final measurement.
● Stop the final measurement.
● Perform a measurement on the current frequency.
4. When the final measurement for the current frequency is done, the R&S
ESW
replaces the scan result in the peak list with the result of the final measurement. If
the frequency has drifted compared to the one of the scan, it also updates the fre-
quency in the peak list.
5. The R&S
ESW continues with the next frequency in the peak list, positions the
marker on that frequency etc.
6. When all frequencies in the peak list are finished, the R&S
ESW opens the "Final
Peak List", a table that contains the results for the final measurement.
Note that it is possible to start with an automatic measurement and later change into
interactive mode. Likewise, it is possible to start measuring in interactive mode and
later change into automatic mode.
...............................................................................................131
Working with the Final Measurement Results
.......................................................133
9.3.5.1
Configuring Traces
Access
: "Overview" > "Test Automation" > "Trace / Final Meas"
Test Automation