![Rohde & Schwarz ESW Series User Manual Download Page 462](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816462.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
462
User Manual 1177.6298.02 ─ 08
Parameters:
<Measurements>
<numeric value> (integer only)
Number of measurements considered in calculating an average
trace.
When you set a count of 1, the R&S
ESW performs a single
measurement over the selected scan range.
Range:
0 to 32767
*RST:
0
Example:
//Define 10 measurements in single measurement mode
SWE:COUN 10
INIT:CONT OFF
//Start sweep and wait for its end
INIT;*WAI
Manual operation:
See
[SENSe:]SWEep:SPACing
<StepMode>
This command selects the frequency step mode.
Note that the command has no effect on the scale and display of the frequency axis.
Parameters:
<StepMode>
LINear
Linear frequency steps with a fix stepsize.
LOGarithmic
Logarithmic frequency steps with the stepsize being a percent-
age of the current frequency.
AUTO
The stepsize is coupled to the resolution bandwidth to get the
best measurement results.
*RST:
AUTO
Example:
//Select logarithmic frequency steps
SWE:SPAC LOG
Manual operation:
See
"Selecting the frequency step mode"
15.5.7
Peak Search
Commands to configure the peak search described elsewhere.
●
CALCulate<n>:MARKer<m>:PEXCursion
.........................................................................463
..........................................................................................463
.................................................................................... 463
....................................................................................... 464
..................................................................................464
CALCulate<n>:PSEarch:CLEar[:IMMediate]
..................................................................... 464
Measurements and Result Displays