Measurements and Result Displays
R&S
®
ESW
130
User Manual 1177.6298.02 ─ 08
Note that the "Comment" column is read-only in the peak list result display. If you
would like to add a comment to a peak, you have to do it in the "Peak List" dialog box.
Modifying the contents of the peak list
When the result display is active, you can modify the peak list directly from within the
result display with the corresponding buttons.
●
Peak Search
Performs a single peak search based on the current scan data.
●
Add to Peak List
➙
"Adding the current frequency to the peak list"
●
Symbols
➙
●
Sort by Freq
and
Sort by Delta
➙
●
Export
and
Separator
➙
9.3.5
Performing Final Measurements
During the final measurement, the R&S
ESW does a measurement on each frequency
in the peak list. When done, it updates the preliminary results in the peak list with those
found during the final measurement. To avoid several tests with different detectors, you
can use several detectors at the same time during the final measurement.
In case you are using a scan table in the test sequence, the final measurement is
based on the contents of the scan table. When the scan table has been
, the
final measurement is based on the current bargraph configuration.
Automatic vs interactive final measurements
The R&S
ESW provides two methods to run a final measurement: an automatic final
measurement and an interactive one.
An automatic final measurement measures all frequencies in the peak list automatically
with limited means of interaction. During an automatic measurement, you can still inter-
rupt ("Hold Final Meas") and resume the measurement or abort it completely ("Stop
Final Meas"). But you will not be able to change the measurement configuration. The
advantage is that you can let the measurement run on its own and do not have to oper-
ate the R&S
ESW.
If you want to be able to control the final measurement, for example to change the con-
figuration during the measurement, the R&S
ESW also provides an interactive final
measurement (which you can select even after you have started an automatic final
measurement).
When you use the interactive final measurement, the R&S
ESW interrupts the mea-
surement before it measures a frequency part of the peak list. In this way, you can cus-
tomize the measurement configuration for each frequency.
Test Automation