Remote Commands in the Receiver Application
R&S
®
ESW
649
User Manual 1177.6298.02 ─ 08
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:ATTen
<State>
This command turns synchronization of the attenuation and unit between measure-
ment channels on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:ATT ALL
Synchronizes the attenuation.
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:BWIDth
<State>
INSTrument:COUPle:BANDwidth
<State>
This command turns synchronization of the resolution bandwidth (and filter type)
between measurement channels on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns on synchronization.
*RST:
NONE
Example:
INST:COUP:BWID ALL
Synchronizes the resolution bandwidth.
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:CENTer
<State>
This command turns synchronization of the frequency between measurement channels
on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
General Instrument Setup