Remote Commands in the Receiver Application
R&S
®
ESW
590
User Manual 1177.6298.02 ─ 08
Parameters:
<arg0>
<arg1>
'Factory'
Restores the factory settings as the default state.
'<file_name>
String containing the path and name of the configuration file.
Note that only
instrument
settings files can be selected for the
startup recall function; channel files cause an error.
Example:
MMEM:LOAD:AUTO 1,'C:\R_S\Instr\User\TEST'
Manual operation:
See
MMEMory:LOAD:STATe
<arg0>, <arg1>
This command restores and activates the <instrument> configuration stored in a *.dfl
file.
Note that files with other formats cannot be loaded with this command.
The contents that are reloaded from the file are defined by the last selection made
either in the "Save/Recall" dialogs (manual operation) or through the
MMEMory:SELect[:ITEM]
commands (remote operation; the settings are identical in
both cases).
By default, the selection is limited to the user settings ("User Settings" selection in the
dialogs,
HWSettings
in SCPI). The selection is not reset by [Preset] or
*RST
.
As a consequence, the results of a SCPI script using the
MMEMory:LOAD:STATe
com-
mand without a previous
MMEMory:SELect[:ITEM]
command may vary, depending
on previous actions in the GUI or in previous scripts, even if the script starts with the
*RST
command.
It is therefore recommended that you use the appropriate
MMEMory:SELect[:ITEM]
command before using
MMEMory:LOAD:STATe
.
Parameters:
<arg0>
<arg1>
String containing the path and name of the file to load.
The string may or may not include the file's extension.
Data Management