Common Measurement Settings
R&S
®
ESW
165
User Manual 1177.6298.02 ─ 08
●
●
Configuring the output of the Video signal ← Selecting the output type
Additional settings for video signal output are not supported.
Configuring the output of an AM signal ← Selecting the output type
For the output of AM demodulated signals, you can adjust the following parameters.
●
●
●
Configuring the output of an FM signal ← Selecting the output type
For the output of FM demodulated signals, you can adjust the following parameters.
●
●
●
●
Controlling and configuring the output
Depending on the selected
, you can configure one or more of the following
output characteristics.
IF Output Frequency ← Controlling and configuring the output
Defines the output frequency of the IF signal.
The range is: (RBW
/
2) to (240
MHz - (RBW
/
2))
Remote command:
Coupling ← Controlling and configuring the output
Selects the type of current that is transferred at the output.
Available for linear signal output.
"AC Coupling"
Rejects the DC component of the signal.
This coupling protects the output from damage, but can distort very
low frequencies.
"DC Coupling"
Transfers the complete signal (DC and AC components).
Remote command:
Reference Level for Output ← Controlling and configuring the output
Shows the reference level of the signal, if the level of the output signal depends on the
reference level of the current measurement.
Remote command:
not supported
Low Pass ← Controlling and configuring the output
Turns a low pass filter to control the frequencies that are output on and off.
Data Input and Output