Remote Commands in the Receiver Application
R&S
®
ESW
503
User Manual 1177.6298.02 ─ 08
Suffix:
<tp>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Usage:
Event
Manual operation:
See
OUTPut:TRIGger<tp>:PULSe:LENGth
<Length>
This command defines the length of the pulse generated at the trigger output.
Suffix:
<tp>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Length>
Pulse length in seconds.
Default unit: S
Example:
OUTP:TRIG2:PULS:LENG 0.02
Manual operation:
See
15.6.3
Frequency Configuration
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe]
....................................................503
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth
.............................................. 504
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold
...............................................504
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe
.................................................... 504
DISPlay[:WINDow<n>]:TRACe<t>:X:SPACing
.................................................................. 505
......................................................................................... 505
................................................................................506
..................................................................................... 506
............................................................................................506
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe]
<State>
This command turns signal tracking on and off.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Configuration