Remote Commands in the Receiver Application
R&S
®
ESW
657
User Manual 1177.6298.02 ─ 08
BIDir
Changes of a parameter get applied both ways (from channel 1
to channel 2 and vice versa).
LTOR
Changes of a parameter get applied from channel 1 to channel
2, but not the other way around.
RTOL
Changes of a parameter get applied from channel 2 to channel
1, but not the other way around.
<State>
ON | OFF
Optional SCPI parameter.
Return values:
<Index>
Index number of the new custom coupling.
Note that the returned index numbers do not necessarily have to
be the same as those shown in the user interface.
Example:
INST:COUP:USER:NEW? 'Receiver','All
Windows','Attenuation','Spectrum','All
Windows','Attenuation',BID,ON
would return, e.g.:
3
Synchronizes the attenuation between the channels named
'Receiver' and 'Spectrum' in both directions and turns on the
coupling. Also returns the index number of the custom coupling.
Example:
INST:COUP:USER:NEW? 'All Receiver','All
Windows','Attenuation','','','',BID,ON
would return, e.g.:
3
Synchronizes the attenuation between all Receiver channels in
both directions and turns on the coupling. Also returns the index
number of the custom coupling.
Usage:
Query only
Manual operation:
See
"Selecting the channels to synchronize"
See
"Selecting the measurement windows to synchronize"
on page 352
See
"Selecting the parameter to synchronize"
See
"Applying the coupling mechanism"
INSTrument:COUPle:USER<j>:NUMBers:LIST?
This command queries the index number of custom synchronization mechanisms.
The index numbers define the range for suffix <j> available for the following com-
mands:
●
●
INSTrument:COUPle:USER<j>:STATe
General Instrument Setup