Remote Commands in the Receiver Application
R&S
®
ESW
456
User Manual 1177.6298.02 ─ 08
[SENSe:]FREQuency:STOP
<Frequency>
This command defines the stop frequency of a scan.
In the Spectrum application, the command defines the stop frequency of a measure-
ment.
Parameters:
<Frequency>
<numeric value>
Range:
Refer to the datasheet
*RST:
depends on application
Default unit: Hz
Example:
//Define stop frequency
FREQ:STOP 100MHz
Manual operation:
See
"Defining a frequency range for the scan"
See
[SENSe:]SCAN<sr>:BANDwidth:RESolution
<Bandwidth>
This command defines the resolution bandwidth applied in the selected scan range.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<Bandwidth>
<numeric value>
If you enter a resolution bandwidth that is not supported, the
R&S
ESW uses the next available bandwidth instead.
Range:
Refer to datasheet
Default unit: Hz
Example:
//Define a measurement bandwidth for the 4th scan range
SCAN4:BAND:RES 1MHz
Manual operation:
See
[SENSe:]SCAN<sr>:BARS
<State>
This command turns the display of the bars indicating the size of a scan range on and
off.
Suffix:
<sr>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON
Example:
//Turn on range bars
SCAN:BARS ON
Manual operation:
See
Measurements and Result Displays