Troubleshooting
R&S
®
ESW
729
User Manual 1177.6298.02 ─ 08
Remote command:
[SENSe:]SWEep:DURation
Multiple user access to one instrument
Using the R&S
ESW's LXI browser interface, several users can access
and operate
the same instrument simultaneously. This is useful for troubleshooting or training pur-
poses.
Type the instrument's host name or IP address in the address field of the browser on
your PC, for example "http://10.113.10.203". The instrument home page (welcome
page) opens.
Note:
This function can be deactivated for the instrument. After a firmware update, it is
automatically activated again.
Web browser access to instrument fails
If an error message ("Failed to connect to server (code. 1006)") is displayed in the web
browser instead of the instrument's user interface then the LXI web browser interface
was probably deactivated.
The transducer factors/limit lines applied to my measurement are different to
those displayed in the Transducer/Lines dialog box
If a transducer file was in use when the save set was stored (with the save item "Cur-
rent Settings" only) it is anticipated that these transducer values should remain valid
after every recall of that save set. Thus, even if the transducer file is changed and the
original save set file is recalled later, the
originally stored
transducer values are recal-
led and applied to the measurement. In the "Transducer" dialog box, however, the
changed
transducer file values are displayed as no updated transducer file was loa-
ded.
The same applies to limit line settings.
If you want to apply the changed transducer values after recalling the save set you
must force the application to reload the transducer file. To do so, simply open the "Edit
Transducer" dialog box (see
Chapter 13.3.2, "Working with Transducers"
and toggle the "X-Axis" option from "lin" to "log" and back. Due to that change, the
transducer file is automatically reloaded, and the changed transducer values are
applied to the current measurement. Now you can create a new save set with the
updated transducer values.
Similarly, if you want to apply the changed limit values after recalling the save set you
must force the application to reload the limit file. To do so, simply open the "Edit Limit
Line" dialog box (see
Chapter 11.5.2.2, "Limit Line Settings and Functions"
on page 245) and toggle the "Y-Axis" unit. Due to that change, the limit line file is auto-
matically reloaded, and the changed limit values are applied to the current measure-
ment. Now a new save set with the updated limit values can be created.
17.5
System Recovery
For instruments running Windows 10
, the system drive is delivered with a recovery
partition that allows you to restore the original operating system image and firmware.
System Recovery