Remote Commands in the Receiver Application
R&S
®
ESW
560
User Manual 1177.6298.02 ─ 08
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum[:PEAK]
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum[:PEAK]
This command moves a delta marker vertically to the minimum level for the current fre-
quency.
The search includes all frames. It does not change the horizontal position of the
marker.
If the marker hasn't been active yet, the command first looks for the peak level in the
whole spectrogram and moves the marker vertically to the minimum level.
Suffix:
<n>
.
<m>
15.7.5
Display and Limit Line Configuration
15.7.5.1
Display Lines
...............................................................................................560
.................................................................................... 561
............................................................................................... 561
.....................................................................................561
..........................................................................................562
............................................................................................... 562
.....................................................................................562
CALCulate<n>:DLINe<dl>
<Position>
This command defines the (horizontal) position of a display line.
Suffix:
<n>
.
<dl>
1 | 2
Parameters:
<Position>
The value range is variable.
You can use any unit you want, the R&S
ESW then converts the
unit to the currently selected unit. If you omit a unit, the
R&S
ESW uses the currently selected unit.
*RST:
(state is OFF)
Default unit: DBM
Example:
CALC:DLIN2 -20dBm
Positions the second display line at -20 dBm.
Manual operation:
See
"Horizontal Line 1/ Horizontal Line 2"
Analysis