Remote Commands in the Receiver Application
R&S
®
ESW
575
User Manual 1177.6298.02 ─ 08
CALCulate<n>:LIMit<li>:UPPer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:THReshold
<Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S
ESW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Limit>
Numeric value.
The unit depends on
on page 573.
*RST:
-200
Default unit: dBm
Manual operation:
See
Limit Check
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
...................................................................... 575
........................................................................................ 576
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on
all
limit lines in
all
measurement windows at the same time.
Suffix:
<n>
.
<li>
irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
Analysis