![Rohde & Schwarz ESW Series User Manual Download Page 351](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816351.webp)
General Instrument Setup
R&S
®
ESW
351
User Manual 1177.6298.02 ─ 08
State
Turns the coupling on and off.
INSTrument:COUPle:USER<j>:STATe
Direction
Selects the direction(s) in which the coupling is applied.
INSTrument:COUPle:USER<j>:RELation
Deletes the selected coupling.
INSTrument:COUPle:USER<j>:REMove
Info
Shows information (for example restrictions) for the selected coupling.
Note that in most cases, no information is displayed.
INSTrument:COUPle:USER<j>:INFO?
Creating and editing synchronization mechanisms
Access
: [SETUP] > "Parameter Coupling" > "User Coupling" > "Add New User Cou-
pling"
User couplings are a way to utilize coupling mechanisms other than those available in
the "General" tab of the coupling manager. Using those allows you to create highly
customized couplings between measurement channels.
Compared to the predefined couplings, user couplings do not necessarily have to syn-
chronize all active measurement channels. Instead you can define specific channels
that are synchronized with each other (in any combination you wish for), while other
channels remain independent.
Example:
You currently run two instances of the spectrum application, two instances of the
receiver application and one instance of the analog demodulator.
In this scenario, you could, for example, synchronize only the first instance of the spec-
trum application with the first instance of the receiver application, while the other three
channels remain independent.
Alternatively, you could, for example, synchronize all instances of the receiver applica-
tion, while the spectrum and analog demodulation applications remain independent.
Selecting the channels to synchronize
........................................................................351
Selecting the measurement windows to synchronize
................................................. 352
Selecting the parameter to synchronize
......................................................................352
Applying the coupling mechanism
...............................................................................353
Selecting the channels to synchronize
You can select the channels you would like to synchronize from the "Channel 1" and
"Channel 2" dropdown menu.
The contents of the dropdown menu depends on the number and type of measurement
channels that are currently active. In addition, the availability of the "Channel 2" drop-
down menu depends on the item you have selected from the "Channel 1" dropdown
menu.
●
<channel name>
Synchronizes
two specific channels
of the same type or of different type.
Synchronizing Measurement Channel Configuration