When you connect the LISN, make sure to use the correct pins on both sides of the
test setup. Otherwise, the automatic phase control of the LISN might not work. The fol-
lowing illustrations show the correct pin assignment.
14
N
N
46
15
L1
L1
21
16
22
17
23
12
GND
GND
50
18
PE fl
PE fl
47
L2
L3
L2
L3
13
+5 V
+5 V
48
ESH2-Z5
Pin
Filter
Adapter Pin
(9-pol
25-pol)
2
6
7
4
8
3
1
R&S Receiver
Pin
Figure 9-3: Connection from R&S
ESW to R&S ESH2-Z5
14
N
N
2
15
L
L
6
18
3
12
8
13
+5 V
+5 V
1
ESH3-Z5
Pin
Filter
PE fl
DGND
PE flo
DGND
Adapter Pin
(9-pol
25-pol)
2
6
3
8
1
R&S Receiver
Pin
Figure 9-4: Connection from R&S
ESW to R&S ESH3-Z5
14
N
N
14
15
L
L
15
18
18
12
12
13
+5 V
+5 V
13
R&S Receiver
Pin
ENV216
Pin
Filter
PE fl
DGND
PE flo
DGND
Adapter Pin
(9-pol
25-pol)
2
6
3
8
1
Figure 9-5: Connection from R&S
ESW to R&S ESH216
14
N
N
14
15
L1
L1
15
16
16
17
17
12
GND
GND
12
13
+5 V
+5 V
13
L2
L3
L2
L3
Filter
ENV4200 / ENV432
Pin
2
6
7
4
8
1
Adapter Pin
(9-pol
25-pol)
R&S Receiver
Pin
Figure 9-6: Connection from R&S
ESW to R&S ENV4200 or R&S ENV432
To control the phase selection and PE simulating network of the V-Networks
R&S
ESH2-Z5, R&S
ESH3-Z5, R&S
ENV4200 or R&S
ENV432, the +5
V supply volt-
age and some control lines have to be routed through the wall of the shielded room.
You can also use a direct connection without a filter, e.g. when you use the R&S
ESW
in a shielded room. In that case, you can use the following cables.
●
R&S ESH2-Z5: EZ-5, EZ-13, EZ14
●
R&S ESH3-Z5: EZ-6, EZ-14
●
R&S ENV216: EZ14
●
R&S ENV4200 or R&S ENV432: EZ-14, EZ-21
Test Automation