
Remote Commands in the Receiver Application
R&S
®
ESW
442
User Manual 1177.6298.02 ─ 08
INITiate<mt>:EMITest
This command initiates an automated test sequence.
The sequence consists of a scan, a peak search and a final measurement.
When you are running a continuous measurement, you have to stop the scan deliber-
ately (with
) before the peak search is performed. For a single measurement,
the R&S
ESW automatically performs the peak search after number of scans defined
by the scan count has been performed.
Suffix:
<mt>
.
irrelevant
Example:
//Start the test sequence
INIT2:EMIT
Usage:
Event
Manual operation:
See
"Performing single measurements"
INITiate<mt>:FMEasurement
This command initiates a final measurement based on the peak list.
Suffix:
<mt>
.
irrelevant
Example:
//Start the final measurement
INIT2:FME
Usage:
Event
Manual operation:
See
"Performing single measurements"
INITiate<mt>[:IMMediate]
The command initiates a new measurement.
For a single measurement, the R&S
ESW stops measuring when it has reached the
end frequency. When you start a continuous measurement, it stops only if you abort it
deliberately.
If you are using trace modes MAXHold, MINHold and AVERage, previous results are
reset when you restart the measurement.
●
Single measurements
Synchronization to the end of the measurement is possible with
*OPC
,
*OPC?
or
*WAI
.
●
Continuous measurements
Synchronization to the end of the measurement is not possible.
It is thus recommended to use a single measurement for remote controlled mea-
surements, because results like trace data or markers are only valid after synchro-
nization.
Measurements and Result Displays