General Instrument Setup
R&S
®
ESW
318
User Manual 1177.6298.02 ─ 08
Transducers in the receiver application
The transducer is calculated for a set scan. The transducer is uniquely calculated for
each frequency point and added to the result of the level measurement as the mea-
surement results are stored internally and can be zoomed later.
Transducers in the spectrum application
The R&S
ESW calculates the correction values defined in the transducer factor for
every displayed measurement point before you start the measurement. During the
measurement, it adds the correction values to the measurement results. If you change
the frequency range of the measurement, the R&S
ESW calculates the correction val-
ues again.
If several measured values are combined, only one value is considered.
13.3.2
Working with Transducers
The R&S
ESW allows you to create or edit transducer factors and transducer sets. The
corresponding functions are combined in a dialog box.
For more information about transducer factors in general, see
Transducer settings in secure user mode
Be sure to store transducer files before
"Protecting Data Using the Secure User Mode"
................................................................................................ 318
...............................................................................................323
13.3.2.1
Using Transducers
Access
: [SETUP] > "Transducer"
Transducers