Remote Commands in the Receiver Application
R&S
®
ESW
625
User Manual 1177.6298.02 ─ 08
[SENSe:]ROSCillator:TRANge
<Range>
Defines the tuning range. The tuning range is only available for the variable external
reference frequency. It determines how far the frequency may deviate from the defined
level in parts per million (10
-6
).
Parameters:
<Range>
WIDE | SMALl
The possible values depend on the reference source (see
SMALl
With this smaller deviation (+/- 0.5 ppm) a very narrow fixed loop
bandwidth of 0.1 Hz is realized. With this setting the instrument
can synchronize to an external reference signal with a very pre-
cise frequency. Due to the very narrow loop bandwidth, unwan-
ted noise or spurious components on the external reference
input signal are strongly attenuated. Furthermore, the loop
requires about 30 seconds to reach a locked state. During this
locking process, "NO REF" is displayed in the status bar.
WIDE
The larger deviation (+/- 6 ppm) allows the instrument to syn-
chronize to less precise external reference input signals.
Example:
ROSC:TRAN WIDE
Manual operation:
See
15.9.3
Calibration and Temperature Check
DIAGnostic:SERVice:INPut:MC[:DISTance]
...................................................................... 626
DIAGnostic:SERVice:INPut:PULSed:CFRequency
............................................................ 627
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
......................................................................627
DIAGnostic:SERVice:INPut[:SELect]
................................................................................ 627
DIAGnostic:SERVice:INPut:SYNThtwo[:FREQuency]
........................................................ 628
DIAGnostic:SERVice:STESt:RESult?
............................................................................... 628
SOURce<si>:TEMPerature:FRONtend
.............................................................................628
CALibration[:ALL]
[<CalState>]
This command initiates a calibration (self-alignment) routine and queries if calibration
was successful.
During the acquisition of correction data the instrument does not accept any remote
control commands.
Note:
If you start a self-alignment remotely, then select the "Local" softkey while the
alignment is still running, the instrument only returns to the manual operation state
after the alignment is completed.
General Instrument Setup