Example:
//Add a frequency to the peak list
CALC:PEAK:ADD 93MHz
Usage:
Setting only
Manual operation:
See
See
"Adding the current frequency to the peak list"
CALCulate<n>:PSEarch:AUTO
<State>
CALCulate<n>:PEAKsearch:AUTO
<State>
This command turns an automatic test sequence on and off.
An automated test sequence includes a scan and a peak search, but no final measure-
ment.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turn on the automated peak search
CALC:PEAK:AUTO ON
CALCulate<n>:PSEarch:CLEar[:IMMediate]
CALCulate<n>:PEAKsearch:CLEar[:IMMediate]
This command deletes the contents of a peak list.
Suffix:
<n>
.
irrelevant
Example:
//Clear the peak list
CALC:PEAK:CLE
Usage:
Event
Manual operation:
See
CALCulate<n>:PSEarch[:IMMediate]
CALCulate<n>:PEAKsearch[:IMMediate]
This command initiates a peak search.
Suffix:
<n>
.
irrelevant
Example:
//Initiate a peak search
CALC:PEAK