![Rohde & Schwarz ESW Series User Manual Download Page 440](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816440.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
440
User Manual 1177.6298.02 ─ 08
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the
*OPC
or
*WAI
command after
ABORt
and
before the next command.
To abort a sequence of measurements by the Sequencer, use the
command.
Note on blocked remote control programs
:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish. The remote chan-
nel to the R&S
ESW is blocked for further commands. In this case, you must interrupt
processing on the remote channel first to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S
ESW on a parallel channel to clear all currently active remote channels. Depend-
ing on the used interface and protocol, send the following commands:
●
Visa
:
viClear()
●
GPIB
:
ibclr()
●
RSIB
:
RSDLLibclr()
Now you can send the
ABORt
command on the remote channel that runs the measure-
ment.
Example:
ABOR;INIT:IMM
Aborts the measurement and restarts it.
Usage:
Event
HOLD
This command interrupts (holds) a scan.
To resume the scan, use
Example:
HOLD
Interrupts the scan.
Usage:
Event
INITiate<mt>:CONMeas
This command resumes a scan that was interrupted by a transducer stop at the current
receiver frequency.
If the scan was interrupted by the
command, you have to resume it with
.
Suffix:
<mt>
.
INITiate1 is irrelevant.
INITiate2 resumes the scan.
Measurements and Result Displays