Overloading
At a reference level of -10 dBm and at an external generator output level of the same
value, the R&S
ESW operates without overrange reserve. That means the R&S
ESW
is in danger of being overloaded if a signal is applied whose amplitude is higher than
the reference line. In this case, either the message "RF OVLD" for overload or "IF
OVLD" for exceeded display range (clipping of the trace at the upper diagram border =
overrange) is displayed in the status line.
Overloading can be avoided as follows:
●
Reducing the output level of the external generator (
in "External Generator > Measurement Configuration")
●
Increasing attenuation in the "Amplitude" menu (
10.6.3.2
External Generator Control Settings
Access
: [INPUT/OUPUT] > "External Generator Config"
The "External Generator" settings are available if the R&S
ESW External Generator
Control option is installed. For each measurement channel, you can configure one
external generator. To switch between different configurations, define multiple mea-
surement channels.
For more information on external generator control, see
Interface Configuration Settings
............................................................................159
.......................................................................................... 161
Interface Configuration Settings
Access
: "Overview" > "Input" > "External Generator" > "Interface Configuration"
GPIB Address/TCPIP Address / Computer Name
...................................................... 160
............................................................................................160
..................................................................................161
Data Input and Output