#41024<value1><value2>…<value n>
with
#4
Number of digits (= 4 in the example) of the following number of data bytes
1024
Number of following data bytes (= 1024 in the example)
<Value>
4-byte floating point value
Reading out data in binary format is quicker than in ASCII format. Thus, binary format
is recommended for large amounts of data.
15.7.4
Markers
Individual Marker Configuration
............................................................................ 537
...............................................................................541
................................................................................................ 545
............................................................................................551
15.7.4.1
Individual Marker Configuration
CALCulate<n>:DELTamarker<m>:AOFF
...........................................................................537
CALCulate<n>:DELTamarker<m>:MODE
......................................................................... 537
CALCulate<n>:DELTamarker<m>:MREFerence
................................................................ 538
CALCulate<n>:DELTamarker<m>[:STATe]
........................................................................ 538
CALCulate<n>:DELTamarker<m>:TRACe
.........................................................................539
................................................................................. 539
..................................................................................539
CALCulate<n>:MARKer<m>[:STATe]
............................................................................... 540
................................................................................540
........................................................................................ 540
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off
all
delta markers.
Suffix:
<n>
.
<m>
irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
CALCulate<n>:DELTamarker<m>:MODE
<Mode>
This command defines whether the position of a delta marker is provided as an abso-
lute value or relative to a reference marker.