Remote Commands in the Receiver Application
R&S
®
ESW
514
User Manual 1177.6298.02 ─ 08
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POSitive
Example:
SWE:EGAT:POL POS
[SENSe:]SWEep:EGATe:SOURce
<TriggerSource>
This command selects a trigger source for gated measurements.
If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth (10 MHz).
Parameters:
<TriggerSource>
EXTernal | EXT2 | EXT3
Selects one of the external trigger connectors.
IMMediate
Free Run mode (= no trigger)
*RST:
IMMediate
Example:
//Turn on a gated measurement using the first trigger input
SWE:EGAT ON
SWE:EGAT:SOUR EXT
Manual operation:
See
[SENSe:]SWEep:EGATe
<State>
This command turns gated measurements on and off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
SWE:EGAT ON
Switches on the gate mode.
SWE:EGAT:TYPE EDGE
Switches on the edge-triggered mode.
SWE:EGAT:HOLD 100US
Sets the gate delay to 100 µs.
SWE:EGAT:LEN 500US
Sets the gate opening time to 500 µs.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
Configuration