![Rohde & Schwarz ESW Series User Manual Download Page 485](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816485.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
485
User Manual 1177.6298.02 ─ 08
INPut<ip>:COUPling
<CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CouplingType>
AC | DC
AC
AC coupling
DC
DC coupling
*RST:
AC
Example:
INP:COUP DC
Manual operation:
See
INPut<ip>:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
INPut<ip>:TYPE
<Input>
The command selects the input path.
Suffix:
<ip>
.
1..n
Parameters:
<Input>
INPUT1
Selects RF input 1.
INPUT2
Selects RF input 2.
*RST:
INPUT1
Example:
//Select input path
INP:TYPE INPUT1
Configuration