Remote Commands in the Receiver Application
R&S
®
ESW
455
User Manual 1177.6298.02 ─ 08
[SENSe:]DETector<t>[:FUNCtion]
<Detector>
This command selects the detector for the scan.
In the Spectrum application, it selects the detector in general.
Suffix:
<t>
.
1..n
Parameters:
<Detector>
AVERage
Selects the Average detector.
CAVerage
Selects the CISPR Average detector.
CRMS
Selects the CISPR RMS detector.
NEGative
Selects the Min Peak detector.
POSitive
Selects the Max Peak detector.
QPEak
Selects the Quasipeak detector.
RMS
Selects the RMS detector.
*RST:
POSitive
Example:
//Select scan detector
DET RMS
Manual operation:
See
[SENSe:]FREQuency:STARt
<Frequency>
This command defines the start frequency of a scan.
In the spectrum application, the command defines the start frequency of a measure-
ment.
Parameters:
<Frequency>
<numeric value>
Range:
Refer to the datasheet
*RST:
depends on application
Default unit: Hz
Example:
//Define start frequency
FREQ:STAR 30 kHz
Manual operation:
See
"Defining a frequency range for the scan"
See
Measurements and Result Displays