Remote Commands in the Receiver Application
R&S
®
ESW
479
User Manual 1177.6298.02 ─ 08
The command returns the following information.
<TraceNumber>,<Detector>,<Frequency>,<Level>,
<DeltaLimit>,<Comment>,...
<DeltaLimit>
is '0' if you use no limit line.
<comment>
is empty if you have not entered one for the corre-
sponding peak.
Example:
//Query peak information
CALC:FME:PLIS4?
//Return value example:
'TRACE2,Average,4.5e+07,53.07,0,Woohoo'
Usage:
Query only
15.5.11
LISN Configuration
[SENSe:]FMEasurement:LISN:PHASe
............................................................................. 479
....................................................................................479
INPut<ip>:LISN:FILTer:HPASs[:STATe]
.............................................................................480
[SENSe:]FMEasurement:LISN:PHASe
<Phase>...
This command selects the LISN phases to be measured in a final measurement.
Parameters:
<Phase>
You can select several phases for the final measurement:
<Phase>,[<Phase>,<Phase>,<Phase>]
L1
L2
Available for networks with four phases (R&S
ESH2Z5,
R&S
ENV4200 and R&S
ENV432)
L3
Available for networks with four phases (R&S
ESH2Z5,
R&S
ENV4200 and R&S
ENV432)
N
*RST:
L1
Example:
//Select phases for the final measurement
FME:LISN:PHAS L1,N
Manual operation:
See
[SENSe:]SCAN<sr>:LISN:PHASe
<Phase>...
This command selects the LISN phases to be measured in a scan.
Measurements and Result Displays