![Rohde & Schwarz ESW Series User Manual Download Page 711](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816711.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
711
User Manual 1177.6298.02 ─ 08
Command
Supported subset
Function
Corresp. HP-
Models
Status
VB
VB <numeric_value>
HZ|KHZ|MHZ|GHZ
VB DN
VB UP
VB AUTO
VB?
Video Bandwidth
HP 856xE/
HP 8566B/
HP 8568B/
HP 8594E
available
VBR
1)
VBR <numeric_value>
VBR DN
VBR UP
VBR?
Video Bandwidth Ratio
HP 856xE/
HP 8566B/
HP 8568B/
HP 8594E
available
VIEW
VIEW TRA|TRB|TRC
HP 856xE/
HP 8566B/
HP 8568B/
HP 8594E
available
VTL
VTL <numeric_value>
DB|DM
VTL DN
VTL UP
VTL?
Video Trigger Level
HP 856xE/
HP 8594E
not available
1) HP 8594E only
2) Command is accepted without error message, but is ignored
15.12.2
Special Features of the Syntax Parsing Algorithms for 8566A and
8568A Models
The command syntax is very different for models A and B. Different names are
assigned to identical instrument functions, and the command structure likewise differs
considerably between models A and models B.
The command structure for models A is as follows:
<command>::= <command
code>[<SPC>][<data>|<step>][<SPC>][<delimiter>][<command
code>]…<delimiter>
<data>::= <Value>[<SPC>][<units
code>][<SPC>][<delimiter>][<SPC>][<data>]…
<step>::= UP|DN
where
<command code>
= see Table "Supported Commands"
<Value>
= integer or floating-point numerical value
<units code>
= DM | -DM | DB | HZ | KZ | MZ | GZ | MV | UV | SC | MS | US
Reference: GPIB Commands of Emulated HP Models