Remote Commands in the Receiver Application
R&S
®
ESW
469
User Manual 1177.6298.02 ─ 08
CALCulate<n>:PSEarch:PLISt<peak>:DELete
CALCulate<n>:PEAKsearch:PLISt<pi>:DELete
This command deletes a peak from the peak list.
Suffix:
<n>
.
irrelevant
<pi>
Example:
//Deletes a peak from the peak list
CALC:PEAK:PLIS4:DEL
Usage:
Event
CALCulate<n>:PSEarch:PLISt<peak>:DELTa?
CALCulate<n>:PEAKsearch:PLISt<pi>:DELTa?
This command queries the distance of a peak to the nearest limit line in the peak
search.
Suffix:
<n>
.
irrelevant
<pi>
Return values:
<Level>
If you are using no limit line, the return value is '0'.
Example:
//Query distance to limit line
CALC:PEAK:PLIS4:DELT?
//would return, e.g.
3.23
Usage:
Query only
CALCulate<n>:PSEarch:PLISt<peak>:DETector?
CALCulate<n>:PEAKsearch:PLISt<pi>:DETector?
This command queries the detector with which a peak was found in the peak search.
Suffix:
<n>
.
irrelevant
<pi>
Return values:
<Detector>
String that contains the name of the detector.
Example:
//Query detector the peak was measured with
CALC:PEAK:PLIS4:DET?
//would return, e.g.
'Average'
Usage:
Query only
Measurements and Result Displays