General Instrument Setup
R&S
®
ESW
352
User Manual 1177.6298.02 ─ 08
The <channel name> of the currently active channels is displayed in the channel
bar. Because channel names are arbitrary, the labels of the items in the dropdown
menu are variable.
When you select a specific channel, the "Channel 2" dropdown menu is available
to select the second channel.
Note
: If you want to synchronize another channel, you have to define an additional
user coupling.
●
All <application type>
Synchronizes all active
channels of the same application
(receiver, spectrum, I/Q
analyzer or analog demodulation).
In that case, the "Channel 2" dropdown menu is not available.
●
All channels
Synchronizes
all active channels
, regardless of their type and name.
This is basically the same as the coupling available in the "General" tab of the cou-
pling manager.
The "Channel 2" dropdown menu is not available for synchronization across all
channels.
Remote command:
Query channel:
INSTrument:COUPle:USER<j>:CHANnel:LIST?
Select channel:
INSTrument:COUPle:USER<j>:NEW?
Select channel:
Selecting the measurement windows to synchronize
Available for synchronization of instances of the analog demodulation application.
When you are synchronizing an instance of the analog demodulation application, you
can also select the measurement window that should be synchronized. This feature is
available, because, in the analog demodulation application, the measurement windows
can have configurations that are independent from each other.
You can select the windows you would like to synchronize from the "Specifics for Win-
dow" dropdown menus.
Remote command:
Query windows:
INSTrument:COUPle:USER<j>:WINDow:LIST?
Select a window with
INSTrument:COUPle:USER<j>:NEW?
Select a window with
Selecting the parameter to synchronize
You can synchronize various parameters and settings from different categories across
measurement channels.
To synchronize a parameter, select one from the "Coupling Elements" list. The list con-
tains only parameters that are available in the application you have selected from the
"Channel <x>" dropdown menu. So, for example, synchronizing the reference level is
not possible if one the channels you would like to synchronize is an instance of the
receiver application.
Note that the "Coupling Elements 2" list is available only when you have selected syn-
chronization across two specific channels (and not all channels of the same kind or all
channels). In addition, the second list shows only parameters that are compatible to
the first channel. If there is no compatible parameter, the list remains empty.
Synchronizing Measurement Channel Configuration