Remote Commands in the Receiver Application
R&S
®
ESW
441
User Manual 1177.6298.02 ─ 08
Example:
//Configure a single measurement that is averaged over 20 mea-
surements
INIT2:CONT OFF
SWE:COUN 20
//Start the measurement and wait for the end of the 20 scans
INIT2;*WAI
//Continue the measurement (next 20 sequences) and wait for
the end
INIT2:CONM;*WAI
Usage:
Event
INITiate<n>:CONTinuous
<State>
This command controls the measurement mode for an individual channel.
Note that in single measurement mode, you can synchronize to the end of the mea-
surement with *OPC, *OPC? or *WAI. In continuous measurement mode, synchroniza-
tion to the end of the measurement is not possible. Thus, it is not recommended that
you use continuous measurement mode in remote control, as results like trace data or
markers are only valid after a single measurement end synchronization.
For details on synchronization see
Chapter 14.1.6, "Command Sequence and Syn-
If the measurement mode is changed for a channel while the Sequencer is active (see
on page 444) the mode is only considered the
next time the measurement in that channel is activated by the Sequencer.
Suffix:
<n>
.
1 | 2
INITiate1 selects single or continuous bargraph measurements.
INITiate2 selects single or continuous scans.
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous measurement
OFF | 0
Single measurement
*RST:
1
Example:
INIT:CONT OFF
Switches the measurement mode to single measurement.
INIT:CONT ON
Switches the measurement mode to continuous measurement.
Manual operation:
See
"Performing continuous measurements"
See
"Performing single measurements"
Measurements and Result Displays