Measurements and Result Displays
R&S
®
ESW
132
User Manual 1177.6298.02 ─ 08
Note that the dialog box to configure traces contains the same functions as the dialog
box to configure the final measurement, but with a slightly different layout.
Figure 9-14: Overview of the Final Meas configuration
The remote commands required to configure the final measurement are described in
Chapter 15.5.9, "Final Measurement (and Trace) Configuration"
Functions in the "Trace / Final Meas" dialog box described elsewhere:
●
Trace Mode
Defines the update mode for subsequent traces.
For more information, see
Chapter 11.3.1.2, "Analyzing Several Traces - Trace Mode"
"Clear Write"
Displays the level that has been measured last at each trace point.
The old trace is overwritten.
"Max Hold"
Displays and saves the highest level that has been measured over
several measurements at each trace point.
"Min Hold"
Displays and saves the lowest level that has been measured over
several measurements at each trace point.
"Average"
Displays and saves the average level that has been determined over
several measurements at each trace point.
The number of measurements depends on the
"View"
Freezes the contents of the trace memory and displays the corre-
sponding trace. The trace is not updated when you start another
measurement.
"Blank"
Removes the trace from the diagram.
Test Automation