General Instrument Setup
R&S
®
ESW
324
User Manual 1177.6298.02 ─ 08
Restoring a previously saved dataset
The "Load File" feature restores a dataset that you have previously created (the data-
set can also be on an external memory device).
After you have restored the dataset, it is added to the table.
"File Explorer": Instead of using the file manager of the R&S
ESW firmware, you can
also use the Microsoft Windows File Explorer to manage files.
Working with the dataset table
The table showing all available datasets contains the most relevant information about a
dataset.
●
Edit a dataset
The pencil icon in the first row opens a dialog box that allows you to change con-
tents of a dataset.
For more information, see
●
Edit the name of a dataset
Changes
of the dataset.
●
Check the connection state
The "State" column shows if an RF switch could be found on the selected RF input
and on the network address defined in the corresponding dataset.
When the RF input = OFF, the LED is always red.
●
Select the RF input
Selects the RF input the measurement is performed on.
●
Define a network address
Selects the
of the RF switch.
●
Delete a dataset
The X icon in the last column deletes the corresponding dataset.
Remote commands:
[SENSe:]CORRection:SWITch:SELect
[SENSe:]CORRection:SWITch:SELect
Restore a dataset:
[SENSe:]CORRection:SWITch:LOAD
Delete a dataset:
[SENSe:]CORRection:SWITch:DELete
Programming a Switch Matrix
Access
: [SETUP] > "Transducer" > "Switch Control" > "Add New"
Access
: [SETUP] > "Transducer" > "Switch Control" > "Edit" (pencil icon)
The RF switch is controlled by the R&S
ESW with a set of remote commands that are
saved in a dataset.
....................................................................................... 325
Connection between R&S ESW to the RF switch
....................................................... 325
Transducers