Remote Commands in the Receiver Application
R&S
®
ESW
614
User Manual 1177.6298.02 ─ 08
Parameters:
<Template>
String containing the name of the template.
Example:
HCOP:TREP:ITEM:TEMP:SAVE 'myTemplate'
Saves a test report template.
Usage:
Event
Manual operation:
See
HCOPy:TREPort:NEW
This command creates a new test report.
Creating a test report deletes all previously saved datasets. The current measurement
results are added as the first dataset to the new report.
The R&S
ESW saves the data selected with
Example:
HCOP:TREP:NEW
Creates a test report.
Usage:
Event
Manual operation:
See
"Adding and removing datasets"
HCOPy:TREPort:PAGesize
<Size>
This command selects the size of the test report document.
Parameters:
<Size>
A4
Document pages have an A4 size.
US
Document pages have a US letter size.
*RST:
A4
Example:
HCOP:TREP:PAG A4
Selects the A4 size for the document.
Manual operation:
See
"General properties of the test report document"
HCOPy:TREPort:PAGecount:STATe
<State>
This command includes or excludes page number from the test report.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
HCOP:TREP:PAG:STAT OFF
Removes page numbers from the test report.
Data Management