General Instrument Setup
R&S
®
ESW
337
User Manual 1177.6298.02 ─ 08
No
device-specific error code
Message
brief description of the message
Component
hardware messages: name of the affected module
software messages: name of the affected software
Date/Time
date and time of the occurrence of the message
Remote command:
13.5.4
Firmware Updates
Access
: [Setup] > "System Configuration" > "Firmware Update"
During instrument start, the installed hardware is checked against the current firmware
version to ensure the hardware is supported. If not, an error message is displayed
("Wrong Firmware Version") and you are asked to update the firmware. Until the firm-
ware version is updated, self-alignment fails. To see which components are not suppor-
ted, see the
.
The firmware on your R&S
ESW may also need to be updated in order to enable addi-
tional new features or if reasons for improvement come up. Ask your sales representa-
tive or check the Rohde&Schwarz website for availability of firmware updates. A firm-
ware update package includes at least a setup file and release notes.
Before updating the firmware on your instrument, read the release notes delivered with
the firmware version.
Installing options in secure user mode
Be sure to perform any firmware updates before
Chapter 4.7, "Protecting Data Using the Secure User Mode"
For restricted users in secure user mode, this function is not available!
System Configuration Settings