General Instrument Setup
R&S
®
ESW
353
User Manual 1177.6298.02 ─ 08
Tip
: The "Category" dropdown menu filters the available parameters by a certain cate-
gory.
Remote command:
Query parameters:
INSTrument:COUPle:USER<j>:ELEMent:LIST?
Select parameter:
INSTrument:COUPle:USER<j>:NEW?
Select parameter:
Applying the coupling mechanism
When you are done configuring a new synchronization mechanism, save your settings
with the "Couple Selected Parameters" button.
The new coupling mechanism is now displayed in the custom coupling table and you
can apply it whenever you like.
Remote command:
INSTrument:COUPle:USER<j>:NEW?
(Both commands also couple the parameters.)
13.7.3
Example for a User-Defined Parameter Coupling
Currently two Spectrum application channels are active, one VSA channel, and two
Analog Demodulation channels.
Synchronizing all Spectrum channels
The following example demonstrates how to synchronize the center frequency in all
Spectrum application channels, while the VSA and Analog Demodulation applications
remain independent.
1. Select the [SETUP] key.
2. Select "Parameter Coupling".
3. Select the "User Coupling" tab.
4. Select "Add New User Coupling".
5. From the "Channel 1" list, select "All Spectrum".
6. From the "Coupling Element 1" list, select "Center Frequency".
7. Select "Couple Selected Parameters".
8. Close the "Parameter Coupling" dialog box.
9. In the first Spectrum channel, change the "Center Frequency" to 1
GHz.
10. Switch to the second Spectrum channel.
The center frequency in the second Spectrum channel is also set to 1
GHz.
Synchronizing Measurement Channel Configuration