
Remote Commands in the Receiver Application
R&S
®
ESW
508
User Manual 1177.6298.02 ─ 08
Example:
//Select low noise auto ranging mode
INP:ATT:AMOD LOWN
Manual operation:
See
INPut<ip>:ATTenuation:AUTO
<State>
This command turns automatic determination of the attenuation level on and off.
When you turn it on, the R&S
ESW selects an attenuation that results in a good signal-
to-noise ratio without overloading the RF input.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
ON | OFF
ON
Receiver: Turns on the auto ranging feature.
OFF
Receiver: Turns off the auto ranging feature.
*RST:
ON
Example:
//Turn on auto ranging
INP:ATT:AUTO ON
Manual operation:
See
See
INPut<ip>:ATTenuation:PROTection[:STATe]
<State>
This command turns the availability of attenuation levels of 10
dB or less on and off.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
1
Example:
//Turn on input protection
INP:ATT:PROT ON
Manual operation:
See
INPut<ip>:GAIN:AUTO
<State>
This command includes and excludes the preamplifier from the auto ranging feature.
Suffix:
<ip>
.
irrelevant
Configuration