Remote Commands in the Receiver Application
R&S
®
ESW
638
User Manual 1177.6298.02 ─ 08
[SENSe:]CORRection:TRANsducer:UNIT
<Unit>
This command selects the unit of the transducer factor.
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Unit>
string as defined in table below
*RST:
DB
Example:
CORR:TRAN:UNIT 'DBUV'
String
Unit
'DB'
dB
'DBM'
dBm
'DBMV'
dBmV
'DBUV'
dB
μ
V
'DBUV/M'
dB
μ
V/m
'DBUA'
dB
μ
A
'DBUA/M'
dB
μ
A/m
'DBPW'
dBpW
'DBPT'
dBpT
[SENSe:]CORRection:TSET:BREak
<State>
This command turns a transducer break on and off.
Before you can use the command, you have to select a transducer set with
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
See
[SENSe:]CORRection:TSET:CATalog?
This command queries all transducer sets stored on the R&S
ESW.
Return values:
<UsedDiskSpace>
Size of all available files containing transducer factors in byte.
<FreeDiskSpace>
Remaining disk space in bytes.
<FileInfo>
String containing the file information.
'<NameFileN>,<SizeFileN>'
<NameFileN>
Name of the file.
General Instrument Setup