R&S
®
ZVT
Connecting External Accessories
1300.0045.62 1.13
E-1
establish a LAN connection proceed as follows:
1. Refer to section
Assigning an IP Address
and learn how to avoid connection errors.
2. Connect an appropriate LAN cable to one of the LAN ports. Use a commercial RJ-45 cable to
establish a non-dedicated network connection, or a cross-over RJ-45 cable to establish a dedicated
connection between the analyzer and a single PC.
Dedicated vs. non-dedicated network connections
There are two methods to establish a LAN connection of the analyzer:
•
A non-dedicated network (Ethernet) connection from the analyzer to an existing network made with
an ordinary RJ-45 network cable. The analyzer is assigned an IP address and can coexist with a
computer and with other hosts on the same network.
•
A dedicated network connection between the analyzer and a single computer made with a cross-over
RJ-45 network cable. The computer must be equipped with a network adapter and is directly
connected to the analyzer. The use of hubs, switches, or gateways is not needed, however, data
transfer is still made using the TCP/IP protocol.
The IP address information is displayed in the
Info – Setup Info
dialog.
Test Setups with Two LAN Connections
The two LAN connectors on the rear panel of the analyzer are equivalent. With one LAN connector
used to establish a connection to a home/company network, the other one can be used to connect an
additional instrument, e.g. an additional analyzer or signal generator:
Defining the network topology: Router vs. network client
With two LAN connections, it is possible to use the analyzer in two alternative ways:
•
As a client participating in two independent networks, one comprising the home network including
the analyzer, the second consisting of the additional test instrument plus the analyzer.
•
As a data router between the additional test instrument and the home network. This configuration
means that the analyzer and the additional test instrument are integrated into a single network.
The network topology is defined in Windows XP's
Control Panel – Network Connections – Local Area
Connection Status – Local Area Connection Properties – Internet Protocol (TCP/IP) Properties –
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......