![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 743](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759743.webp)
R&S
®
ZVA/ZVB/ZVT
SOURce<Ch>:POWer...
1145.1084.12 6.281
E-6
SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate]:OFFSet <numeric_value>, ONLY |
CPADd
Defines a port-specific source power or a power offset relative to the channel power
(
SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:AMPlitude]
). An additional
Cal Power
Offset
can be defined via
SOURce<Ch>:POWer<Pt>:CORRection:LEVel:OFFSet
.
<Ch>
Channel number.
<Pt>
Test port number of the analyzer.
<numeric_value>
Port-specific power offset.
Range [def. unit]
–300 dB to +300 dB [dB]. UP and DOWN increment/decrement
the phase in 1-dB steps.
*RST value
0 dB
Offset parameters
ONLY – only the port-specific power is used; the channel power
is ignored.
CPADd – the port-specific power is added as an offset to the
channel power.
*RST value
CPADd
SCPI, Command
Types
Confirmed (with additional device-specific character data
parameters), command or query
Example:
*RST; SOUR:POW -6
Reset the instrument, activating a frequency sweep and set the
internal source power (channel power) for the default channel 1
to -6 dBm.
SOUR:POW1:OFFS 6, ONLY; SOUR:POW2:OFFS 6, CPAD
Replace the source power at port 1 by +6 dBm, the source power
at port 2 by 0 dBm. The powers at the remaining ports (if
available) are not affected.
SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate]:SLOPe <numeric_value>
Defines a linear factor to modify the internal source power at port <Pt> as a function of the stimulus
frequency. The value can be set for frequency sweeps only.
<Ch>
Channel number.
<Pt>
Test port number of the analyzer.
<numeric_value>
Port-specific slope factor.
Range [def. unit]
–40 dB/GHz to +40 dB/GHz. The resulting power range over the
entire frequency sweep must be within the power range of the
analyzer; refer to the data sheet [dBm]. UP and DOWN
increment/decrement the source power in 0.1-dB/GHz steps.
*RST value
0 dB/GHz
SCPI, Command
Types
Device-specific, command or query
Example:
See SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:AMPlitude]
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......