![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 396](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759396.webp)
Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.234
E-6
The
Mechanical Length
dialog contains the three editable columns
Mech. Length, Permittivity
(
r
) and
Velocity Fact.
The velocity factor is 1/sqrt(
r
) and is a measure for the velocity of light in a dielectric with
permittivity
r
relative to the velocity of light in the vacuum (velocity factor < 1); see
Offset parameters:
Definition
above. Permittivity and velocity factor overwrite each other.
Click
Same Dielectric at Each Port
to change the permittivity or velocity factor for all port by
entering a single value.
Remote
control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:DISTance <numeric
_value>
[SENSe<Ch>:]CORRection:EDELay<port_no>:DIELectric <numeric
_value>
Delay
Opens a dialog to define the offset parameters for the different physical and logical test ports as delays.
The delay is calculated as the product of the mechanical length and the square root of the
permittivity
r
divided by the velocity of light in the vacuum; see
Offset parameters: Definition
above.
Open the
Mechanical Length
dialog to change the permittivity.
Remote
control:
[SENSe<Ch>:]CORRection:EDELay<port_no>[:TIME] <numeric
_value>
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......